东芝CT机spellman电源维修CT42 CT70北京
SPELLMAN高压电源广泛应用于东芝CT机XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下
Cause of problem
Object part
24kW
48kW/60kW
E13
Arc-
Cathode side Arc of the X-ray tube or the high-voltage section
TUBE, INV/HV
TUBE, INV/HVC
E25
Arc+
Anode side Arc of the X-ray tube or the high-voltage section
TUBE, INV/HVA
E12
Over kV-
Cathode side overvoltage. Abnormal main circuit control.
INV/HV, SCB
INV/HVC, SCB
E27
Over kV+
Anode side overvoltage. Abnormal main circuit control.
INV/HVA, SCB
E62
Over mA-
Cathode side overcurrent of the X-ray tube or the high-voltage section.
TUBE, INV/HV, SCB
TUBE, INV/HVC,SCB
E63
Over mA+
Anode side overcurrent of the X-ray tube or the high-voltage section.
TUBE, INV/HVA,SCB
E66
MAC error
Filament current abnormal. Trouble of the filament heatig circuit.
INV/HV, SCB
INV/HVC, SCB
E0
Regulation error (HV abnormal)
There is a difference in binding output and real output
INV/HV, AC/C
INV/HVC, INV/HVA, AC/C